SAC-SINGLAS Accredited · ISO/IEC 17025

Thermocouple Calibration in Singapore

Unitest calibrates thermocouples in Singapore across -95 °C to 1000 °C (Types K, J, T, N, R, S, E), under SAC-SINGLAS accreditation LA-2023-0845-C (ISO/IEC 17025). Standard turnaround is 5–7 working days; express service is 2–3 days, with free collection and delivery in Singapore for orders above $300.

Also known as: Thermocouple Sensor · Standard Thermocouple · Type K Thermocouple

Why calibrate?

Unitest provides SAC-SINGLAS accredited thermocouple calibration in Singapore for Types K, J, T, N, R, S and E over a range of -95°C to 1000°C, with traceable ISO/IEC 17025 certificates — in our lab or on-site.

What you receive

What thermocouple calibration at Unitest covers

Traceable certificate

Calibration certificate with an unbroken traceability chain to national and international standards, recording the as-found condition before any adjustment.

Stated uncertainty

Results reported with measurement uncertainty, so you know exactly how much confidence each reading carries in your quality system.

5–7 day turnaround

Standard turnaround 5–7 working days; express 2–3 working days on request. Free collection and delivery within Singapore for orders above S$300.

Part of our temperature calibration service, accredited under SAC-SINGLAS LA-2023-0845-C.

How often

How often should a thermocouple be calibrated?

Thermocouples are typically calibrated every 12 months. Thermocouples in high-temperature or chemically aggressive environments may need 6-monthly checks.

Accredited ranges

Thermocouple calibration ranges & measurement uncertainty

The parameters, ranges and calibration & measurement capability (CMC) below are taken from our SAC-SINGLAS Schedule of Accreditation (LA-2023-0845-C, to ISO/IEC 17025). CMC is the smallest measurement uncertainty we can achieve under accredited conditions.

Thermocouple Wire with Indicator (Lab)· UNI-T001 Rev:0
Range / measured quantityMeasurement uncertainty (CMC)
-80 °C - -40 °C0.17 °C
>-40 °C – 0 °C0.17 °C
>0 °C – 50 °C0.17 °C
>50 °C – 140 °C0.17 °C
>140 °C - -210 °C0.17 °C
>210 °C – 270 °C0.17 °C
>270 °C – 450 °C0.17 °C
>450 °C – 500 °C0.17 °C
>500 °C – 660 °C0.53 °C
>660 °C – 800 °C0.9 °C
>800 °C – 1000 °C0.9 °C
Thermocouple Wire with Indicator (Site)· UNI-T001 Rev:0
Range / measured quantityMeasurement uncertainty (CMC)
-95 °C – 0 °C0.29 °C
>0 °C – 60 °C0.24 °C
>60 °C – 140 °C0.24 °C
>140 °C – 270 °C0.31 °C
>270 °C – 450 °C0.33 °C
>450 °C – 660 °C0.4 °C
Thermocouple Wire without Indicator (Lab)· UNI-T004 Rev:0
Range / measured quantityMeasurement uncertainty (CMC)
-80 °C - -40 °C0.8 °C
>-40 °C – 0 °C0.8 °C
>0 °C – 50 °C0.36 °C
>50 °C – 140 °C0.36 °C
>140 °C – 210 °C0.36 °C
>210 °C – 270 °C0.36 °C
>270 °C – 450 °C0.36 °C
>450 °C – 500 °C0.36°C
>500 °C – 660 °C0.5 °C
>660 °C – 800 °C1.0 °C
>800 °C – 1000 °C1.1 °C
Thermocouple Wire without Indicator (Site)· UNI-T004 Rev:0
Range / measured quantityMeasurement uncertainty (CMC)
-95 °C – 60 °C0.72 °C
>60 °C – 140 °C0.71 °C
>140 °C – 280 °C0.77 °C
>280 °C – 450 °C0.78 °C
>450 °C – 660 °C0.8 °C
Download the full SINGLAS Schedule (LA-2023-0845-C) — PDF

You can also verify our accreditation on the SAC-SINGLAS directory.

Questions

Thermocouple calibration FAQ

How often should a thermocouple be calibrated?

Thermocouples are typically calibrated every 12 months. Thermocouples in high-temperature or chemically aggressive environments may need 6-monthly checks.

Is thermocouple calibration accredited?

Yes. Thermocouple calibration is covered under our SAC-SINGLAS scope of accreditation (LA-2023-0845-C, ISO/IEC 17025), covering -95 °C to 1000 °C (Types K, J, T, N, R, S, E). Certificates state the measured values and measurement uncertainty, traceable to national and international standards. Where an instrument falls outside the accredited ranges, we can calibrate it as a traceable service instead.

How long does thermocouple calibration take?

Standard turnaround is 5 to 7 working days; express service is 2 to 3 working days. Collection and delivery in Singapore are free for orders above $300.

Can you calibrate my brand of thermocouple?

Almost certainly — the makes listed on this site are simply the ones customers most commonly send in. Send us the make and model and our engineers will confirm whether it falls within our accredited scope or can be calibrated as a traceable service.

Related

Related instruments

Browse every instrument type on the Find Your Instrument page.

Need thermocouple calibration?

Submit your instrument list — we'll confirm capability, turnaround and pricing, no obligation.