Platinum Resistance Thermometer (PRT) Calibration in Singapore
Unitest calibrates platinum resistance thermometers (PRTs) in Singapore across -95 °C to 660 °C, plus fixed points from -196 °C to 660.3226 °C, under SAC-SINGLAS accreditation LA-2023-0845-C (ISO/IEC 17025). Standard turnaround is 5–7 working days; express service is 2–3 days, with free collection and delivery in Singapore for orders above $300.
Also known as: Platinum Resistance Thermometer Probe · Precision RTD Thermometer · SPRT
Why calibrate?
Unitest calibrates platinum resistance thermometers (PRTs) and standard platinum resistance thermometers (SPRTs) in Singapore over a range of -95°C to 660°C, with SAC-SINGLAS accredited ISO/IEC 17025 certificates.
What you receive
What platinum resistance thermometer (PRT) calibration at Unitest covers
Traceable certificate
Calibration certificate with an unbroken traceability chain to national and international standards, recording the as-found condition before any adjustment.
Stated uncertainty
Results reported with measurement uncertainty, so you know exactly how much confidence each reading carries in your quality system.
5–7 day turnaround
Standard turnaround 5–7 working days; express 2–3 working days on request. Free collection and delivery within Singapore for orders above S$300.
Part of our temperature calibration service, accredited under SAC-SINGLAS LA-2023-0845-C.
How often
How often should a platinum resistance thermometer (PRT) be calibrated?
PRTs used as working standards should be calibrated every 12 months, or whenever repaired or suspected of damage.
Accredited ranges
Platinum Resistance Thermometer (PRT) calibration ranges & measurement uncertainty
The parameters, ranges and calibration & measurement capability (CMC) below are taken from our SAC-SINGLAS Schedule of Accreditation (LA-2023-0845-C, to ISO/IEC 17025). CMC is the smallest measurement uncertainty we can achieve under accredited conditions.
RTD, PRT Sensor with Indicator & Temperature Measuring instruments (Lab)· UNI-T001 Rev:0
| Range / measured quantity | Measurement uncertainty (CMC) |
|---|---|
| -80 °C – -40 °C | 0.011 °C |
| >-40 °C – 0 °C | 0.01 °C |
| >0 °C – 50 °C | 0.01 °C |
| >50 °C – 140 °C | 0.01 °C |
| >140 °C – 210 °C | 0.01 °C |
| >210 °C – 270 °C | 0.011 °C |
| >270 °C – 450 °C | 0.018 °C |
| >450 °C – 500 °C | 0.02 °C |
| >500 °C – 660 °C | 0.39 °C |
RTD, PRT Sensor with Indicator & Temperature Measuring instruments (Site)· UNI-T001 Rev:0
| Range / measured quantity | Measurement uncertainty (CMC) |
|---|---|
| -95 °C – 0 °C | 0.066 °C |
| >0 °C – 60 °C | 0.033 °C |
| >60 °C – 140 °C | 0.047 °C |
| >140 °C – 270 °C | 0.26 °C |
| >270 °C – 450 °C | 0.33 °C |
| >450 °C – 660 ° | 0.39 °C |
RTD Sensor without Indicator, PRT Sensor without Indicator & Temperature measuring instruments (Lab)· UNI-T003 Rev:0
| Range / measured quantity | Measurement uncertainty (CMC) |
|---|---|
| -80 °C - -40 °C | 0.01 °C |
| >-40 °C – 0°C | 0.01 °C |
| >0 °C – 50 °C | 0.01 °C |
| >50 °C – 140 °C | 0.012 °C |
| >140 °C – 210 °C | 0.012 °C |
| >210 °C – 270 °C | 0.02 °C |
| >270 °C – 500 °C | 0.03 °C |
| >500 °C – 660 °C | 0.39 °C |
RTD Sensor without Indicator, PRT Sensor without Indicator & Temperature measuring instruments (Site)· UNI-T003 Rev:0
| Range / measured quantity | Measurement uncertainty (CMC) |
|---|---|
| -95 °C – 60 °C | 0.6 °C |
| >60 °C – 140 °C | 0.35 °C |
| >140 °C – 270 °C | 0.43 °C |
| >270 °C – 450 °C | 0.47 °C |
| >450 °C – 660 °C | 0.52 °C |
Fixed Point RTD Sensor· UNI-T002 Rev:0
| Range / measured quantity | Measurement uncertainty (CMC) |
|---|---|
| Approx. -196.0 °C | 0.012 °C |
| 0.01 (TPW) | 3 mK |
| 29.7647 (Gallium) | 3 mK |
| 660.3226 (Aluminium) | 9 mK |
You can also verify our accreditation on the SAC-SINGLAS directory.
Questions
Platinum Resistance Thermometer (PRT) calibration FAQ
How often should a platinum resistance thermometer (PRT) be calibrated?
PRTs used as working standards should be calibrated every 12 months, or whenever repaired or suspected of damage.
Is platinum resistance thermometer (PRT) calibration accredited?
Yes. Platinum Resistance Thermometer (PRT) calibration is covered under our SAC-SINGLAS scope of accreditation (LA-2023-0845-C, ISO/IEC 17025), covering -95 °C to 660 °C, plus fixed points from -196 °C to 660.3226 °C. Certificates state the measured values and measurement uncertainty, traceable to national and international standards. Where an instrument falls outside the accredited ranges, we can calibrate it as a traceable service instead.
How long does platinum resistance thermometer (PRT) calibration take?
Standard turnaround is 5 to 7 working days; express service is 2 to 3 working days. Collection and delivery in Singapore are free for orders above $300.
Can you calibrate my brand of platinum resistance thermometer (PRT)?
Almost certainly — the makes listed on this site are simply the ones customers most commonly send in. Send us the make and model and our engineers will confirm whether it falls within our accredited scope or can be calibrated as a traceable service.
Related
Related instruments
Thermocouple
Thermocouple calibration →RTD / PT100 Temperature Sensor
RTD / PT100 Temperature Sensor calibration →Thermistor Temperature Sensor
Thermistor Temperature Sensor calibration →Browse every instrument type on the Find Your Instrument page.