Temperature Signal Transmitter Calibration in Singapore

Unitest calibrates temperature signal transmitters in Singapore across RTD simulation -200 °C to 800 °C, thermocouple simulation -250 °C to 1770 °C (type-dependent) and 4–20 mA output measurement, under SAC-SINGLAS accreditation LA-2023-0845-C (ISO/IEC 17025). Standard turnaround is 5–7 working days; express service is 2–3 days, with free collection and delivery in Singapore for orders above $300.

Also known as: RTD Transmitter · Thermocouple Transmitter

Why calibrate?

Unitest calibrates temperature signal transmitters in Singapore — converting RTD or thermocouple inputs to a 4–20 mA or digital output — with traceable ISO/IEC 17025 certificates. Transmitter calibration verifies both the sensor input response and the signal output.

What you receive

What temperature signal transmitter calibration at Unitest covers

Traceable certificate

Calibration certificate with an unbroken traceability chain to national and international standards, recording the as-found condition before any adjustment.

Stated uncertainty

Results reported with measurement uncertainty, so you know exactly how much confidence each reading carries in your quality system.

5–7 day turnaround

Standard turnaround 5–7 working days; express 2–3 working days on request. Free collection and delivery within Singapore for orders above S$300.

Part of our temperature calibration service, accredited under SAC-SINGLAS LA-2023-0845-C.

How often

How often should a temperature signal transmitter be calibrated?

Temperature transmitters are typically calibrated every 12 months, or more frequently in safety-critical loops.

Accredited ranges

Temperature Signal Transmitter calibration ranges & measurement uncertainty

The parameters, ranges and calibration & measurement capability (CMC) below are taken from our SAC-SINGLAS Schedule of Accreditation (LA-2023-0845-C, to ISO/IEC 17025). CMC is the smallest measurement uncertainty we can achieve under accredited conditions.

RTD Simulator – TYPE: PT 385 (100Ω) (Lab / Site)· UNI-T006 Rev:0
Range / measured quantityMeasurement uncertainty (CMC)
-200 °C – 80 °CLab 0.02 °C / Site 0.15 °C
>80 °C – 100 °CLab 0.01 °C / Site 0.15 °C
>100 °C – 200 °CLab 0.01 °C / Site 0.16 °C
>200 °C – 400 °CLab 0.01 °C / Site 0.27 °C
>400 °C – 800 °CLab 0.01 °C / Site 0.27 °C
Thermocouple Simulator – Type K (Lab / Site)· UNI-T005 Rev:0
Range / measured quantityMeasurement uncertainty (CMC)
-200 °C - -100 °CLab 0.51 °C / Site 0.62 °C
>-100 °C - -25 °CLab 0.42 °C / Site 0.55 °C
>25 °C – 120 °CLab 0.46 °C / Site 0.55 °C
>120 °C – 600 °CLab 0.46 °C / Site 0.55 °C
>600 °C – 1000 °CLab 0.56 °C / Site 0.55 °C
>1000 °C – 1372 °CLab 0.56 °C / Site 0.55 °C
DC Current measuring instruments· UNI-E001 Rev:0
Range / measured quantityMeasurement uncertainty (CMC)
0 µA – 120 µA0.021 µA
>120 µA – 1.2 mA0.045 %
>1.2 mA – 12 mA0.019 %
>12 mA – 120 mA0.02 %
>120 mA – 1.2 A0.029 %
>1.2 A – 3.1 A0.044 %
>3.1 A – 12 A0.041 %
>12 A – 30 A0.11 %
Download the full SINGLAS Schedule (LA-2023-0845-C) — PDF

You can also verify our accreditation on the SAC-SINGLAS directory.

Questions

Temperature Signal Transmitter calibration FAQ

How often should a temperature signal transmitter be calibrated?

Temperature transmitters are typically calibrated every 12 months, or more frequently in safety-critical loops.

Is temperature signal transmitter calibration accredited?

Yes. Temperature Signal Transmitter calibration is covered under our SAC-SINGLAS scope of accreditation (LA-2023-0845-C, ISO/IEC 17025), covering RTD simulation -200 °C to 800 °C, thermocouple simulation -250 °C to 1770 °C (type-dependent) and 4–20 mA output measurement. Certificates state the measured values and measurement uncertainty, traceable to national and international standards. Where an instrument falls outside the accredited ranges, we can calibrate it as a traceable service instead.

How long does temperature signal transmitter calibration take?

Standard turnaround is 5 to 7 working days; express service is 2 to 3 working days. Collection and delivery in Singapore are free for orders above $300.

Can you calibrate my brand of temperature signal transmitter?

Almost certainly — the makes listed on this site are simply the ones customers most commonly send in. Send us the make and model and our engineers will confirm whether it falls within our accredited scope or can be calibrated as a traceable service.

Related

Related instruments

Browse every instrument type on the Find Your Instrument page.

Need temperature signal transmitter calibration?

Submit your instrument list — we'll confirm capability, turnaround and pricing, no obligation.

Get a QuoteWhatsApp