SAC-SINGLAS Accredited · ISO/IEC 17025

Electrical Installation Tester Calibration in Singapore

Unitest calibrates multifunction electrical installation testers in Singapore — including Fluke and Kyoritsu — across insulation 10 kΩ to 999.9 GΩ, earth resistance 100 mΩ to 10 kΩ and RCD trip current 3 mA to 3000 mA, under SAC-SINGLAS accreditation LA-2023-0845-C (ISO/IEC 17025). Standard turnaround is 5–7 working days; express service is 2–3 days, with free collection and delivery in Singapore for orders above $300.

Also known as: Installation Multifunction Tester · RCD Tester

Why calibrate?

Unitest calibrates multifunction electrical installation testers in Singapore — verifying insulation resistance, earth continuity, loop impedance and RCD trip-time functions — with traceable ISO/IEC 17025 certificates. Installation testers used for periodic inspection under SS CP 38 must carry a current calibration certificate.

What you receive

What electrical installation tester calibration at Unitest covers

Traceable certificate

Calibration certificate with an unbroken traceability chain to national and international standards, recording the as-found condition before any adjustment.

Stated uncertainty

Results reported with measurement uncertainty, so you know exactly how much confidence each reading carries in your quality system.

5–7 day turnaround

Standard turnaround 5–7 working days; express 2–3 working days on request. Free collection and delivery within Singapore for orders above S$300.

Part of our electrical calibration service, accredited under SAC-SINGLAS LA-2023-0845-C.

How often

How often should a electrical installation tester be calibrated?

Electrical installation testers should be calibrated every 12 months.

Accredited ranges

Electrical Installation Tester calibration ranges & measurement uncertainty

The parameters, ranges and calibration & measurement capability (CMC) below are taken from our SAC-SINGLAS Schedule of Accreditation (LA-2023-0845-C, to ISO/IEC 17025). CMC is the smallest measurement uncertainty we can achieve under accredited conditions.

High Resistance (Insulation resistance) – measuring instruments· UNI-E004 Rev:0
Range / measured quantityMeasurement uncertainty (CMC)
10 kΩ – 19.999 kΩ0.025 kΩ
20 kΩ – 39.99 kΩ0.047 kΩ
40 kΩ – 99.99 kΩ0.12 kΩ
100 kΩ – 199.99 kΩ0.25 kΩ
200 kΩ – 999.9 kΩ0.47 kΩ
1 MΩ – 1.999 MΩ0.0081 MΩ
2 MΩ – 9.999 MΩ0.01 MΩ
10 MΩ – 19.999 MΩ0.092 MΩ
20 MΩ – 199.99 MΩ0.60 MΩ
200 MΩ – 999.9 MΩ1.4 MΩ
1 GΩ – 1.9 GΩ0.074 GΩ
2 GΩ – 10 GΩ0.077 GΩ
100 GΩ0.15 GΩ
Low Resistance (Earth resistance) – measuring instruments· UNI-E004 Rev:0
Range / measured quantityMeasurement uncertainty (CMC)
100 mΩ – 0.199 Ω0.013 Ω
0.2 Ω – 0.499 Ω0.013 Ω
0.5 Ω – 1.999 Ω0.051 Ω
2 Ω – 4.99 Ω0.019 Ω
5 Ω – 29.9 Ω0.026 Ω
30 Ω – 199.9 Ω0.12 Ω
200 Ω – 499.99 Ω0.49 Ω
500 Ω – 1.999 kΩ0.0014 kΩ
2 kΩ – 4.99 kΩ0.005 kΩ
5 kΩ – 10 kΩ0.014 kΩ
Residual Current measuring instruments· UNI-E004 Rev:0
Range / measured quantityMeasurement uncertainty (CMC)
Trip current: 3 mA – 3000 mA1.8 %
Download the full SINGLAS Schedule (LA-2023-0845-C) — PDF

You can also verify our accreditation on the SAC-SINGLAS directory.

Brands & models

Electrical Installation Tester brands & models we calibrate

The models customers most commonly send in — not our full capability. Every model is quoted individually; request a quotation and we'll confirm ranges, points and turnaround.

BrandModel
Fluke9040
Fluke1663
Fluke1664FC
Kyoritsu6010B

Don't see your model? Send us the make and model — or WhatsApp us a photo of the nameplate — and we'll confirm capability and quote.

Questions

Electrical Installation Tester calibration FAQ

How often should a electrical installation tester be calibrated?

Electrical installation testers should be calibrated every 12 months.

Is electrical installation tester calibration accredited?

Yes. Electrical Installation Tester calibration is covered under our SAC-SINGLAS scope of accreditation (LA-2023-0845-C, ISO/IEC 17025), covering insulation 10 kΩ to 999.9 GΩ, earth resistance 100 mΩ to 10 kΩ and RCD trip current 3 mA to 3000 mA. Certificates state the measured values and measurement uncertainty, traceable to national and international standards. Where an instrument falls outside the accredited ranges, we can calibrate it as a traceable service instead.

How long does electrical installation tester calibration take?

Standard turnaround is 5 to 7 working days; express service is 2 to 3 working days. Collection and delivery in Singapore are free for orders above $300.

Which brands of electrical installation tester do you calibrate?

Commonly submitted brands include Fluke and Kyoritsu. This list covers the models customers most often send in, not our full capability — send us your make and model and we'll confirm whether it falls within our accredited scope or can be calibrated as a traceable service.

Related

Related instruments

Browse every instrument type on the Find Your Instrument page.

Need electrical installation tester calibration?

Submit your instrument list — we'll confirm capability, turnaround and pricing, no obligation.

Get a QuoteWhatsApp